E1 temperature sensor fault requires PT100 replacement; test resistance (100 ohms at 0°C, 138 ohms at 100°C) before ordering to confirm sensor failure.
Tuttnauer autoclave fault codes for 2340, 3870, Elara, and Valueklave series: error codes, causes, and troubleshooting steps.
Parts to price first
E1 temperature sensor fault requires PT100 replacement; test resistance (100 ohms at 0°C, 138 ohms at 100°C) before ordering to confirm sensor failure.
E3 and E5 door faults trace to worn gaskets preventing proper seal; replace if gasket shows cracks, tears, or prevents door from locking under pressure.
E8 conductivity faults and E7 water level issues require distilled or RO water below 15 µS/cm; tap water causes mineral scaling and damages heating elements.
E3 slow heat-up faults trace to scaled or failed heating elements; descale first, replace element if cycle time remains excessive after cleaning.
E4 overpressure faults indicate safety valve failure or blockage; replace valve if manual test shows it does not open at rated pressure.
Limescale buildup causes E3 temperature faults and slow cycles; descale quarterly in hard water areas to maintain heating element efficiency.
The full OEM-style workshop manual has the wiring diagram, test values, and the fault table this guide summarizes. Instant PDF download.
Tuttnauer autoclave fault codes for 2340, 3870, Elara, and Valueklave series: error codes, causes, and troubleshooting steps.
Tuttnauer autoclaves (2340, 3870, Elara, Valueklave, and EZ10 series) display fault codes on the digital or LED display panel. Common in dental, veterinary, and laboratory settings.
| Code | Model | Meaning | Quick Fix |
|---|---|---|---|
| E1 | All | Temperature sensor fault | Check sensor wiring and resistance |
| E2 | All | Pressure sensor fault | Check transducer signal |
| E3 | 2340/3870 | Temperature not reached in time | Check steam supply and door seal |
| E4 | All | Overpressure safety fault | Check safety valve |
| E5 | All | Door not closed or locked | Check door and locking mechanism |
| E6 | Electronic | Memory/EEPROM fault | Contact service |
| E7 | Elara | Water level fault | Check water reservoir |
| E8 | Elara | Conductivity fault | Check water quality (use distilled) |
| E9 | EZ10 | Printer fault | Check printer paper/connection |
| LO | Display | Low water level | Refill with distilled water |
Tuttnauer uses PT100 temperature sensors in the chamber. An E1 fault means the sensor is reading out of range, either a broken wire (reads too high) or a short (reads too low). Disconnect the sensor and test resistance: 100 ohms at 0°C, approximately 138 ohms at 100°C. Replace the sensor if values are incorrect.
If the autoclave doesn’t reach the target temperature (121°C or 134°C) within the preset time, E3 aborts the cycle. This is commonly caused by: door not sealing fully (check gasket), low steam generation from a scaled heating element, or insufficient water. Descale the unit if limescale buildup is visible.
Tuttnauer door faults indicate the door locking mechanism didn’t complete. On manual-lock models, ensure the door handle is fully rotated to the locked position. On automatic-lock models, check the door solenoid and locking cam. A worn door gasket that lets steam escape also prevents the door from locking under pressure.
Elara and EZ10 models have internal water reservoirs with level sensors. Always use distilled or RO water. Tap water causes mineral scaling that damages the heating element and clogs the strainer. Clean the water reservoir and level sensor if false water level alarms occur with a full reservoir.
Elara models test water conductivity to prevent mineral buildup damage. If tap water or inadequately treated water is used, E8 triggers and the cycle is aborted. Use only distilled or deionized water with conductivity below 15 µS/cm.
| Part | Notes |
|---|---|
| PT100 temperature sensor | Replace on E1 |
| Door gasket (silicone) | Replace on E5 / poor sealing |
| Heating element | Replace on slow heat-up |
| Safety valve | Replace on overpressure fault |
| Level sensor | Replace on false LO alarm |
Tuttnauer autoclave spore testing, cycle validation, and annual PM require trained biomedical technicians. In dental and medical practices, sterilizer validation records are required by regulatory bodies.
E1 indicates a PT100 temperature sensor fault from broken wiring or sensor failure. Disconnect the sensor and test resistance with a multimeter (100 ohms at 0°C). Replace the sensor if values are incorrect or if you see an open circuit.
E3 occurs when the unit cannot reach 121°C or 134°C in the preset time. Check door gasket seal, descale the heating element if limescale is visible, verify water level, and ensure steam generation is adequate.
E8 triggers when water conductivity exceeds 15 µS/cm. Use only distilled or deionized water. Drain and refill the reservoir with proper water to clear the fault.
Call for spore testing, cycle validation, annual preventive maintenance, or if faults persist after basic troubleshooting. Medical and dental practices require documented validation records for regulatory compliance.